Melt Flow Index Tester: MFI Formula, Test Procedure and Calculation Example

A Melt Flow Index Tester measures how quickly a molten thermoplastic passes through a specified die under a defined temperature and applied load. For the mass-based test, the MFI formula is MFR = 600 × m ÷ t, with m in grams and t in seconds. If a valid cut weighs 0.365 g and represents 30.0 s of flow, the calculated result is 7.30 g/10 min. The 600 converts the observed rate to a ten-minute basis. The actual measurement interval follows the applicable method and material requirements.

That number is usable only with its test conditions. Check the standard, edition, procedure, temperature, load and die before comparing lots, transferring a method or auditing an instrument result. This guide explains the direct mass calculation, the displacement calculation for volume flow and when a conversion using melt density is valid. The examples teach the arithmetic. Use your controlled procedure and instrument instructions for actual settings, safe operation and acceptance criteria.

What does a Melt Flow Index Tester measure?

The instrument is an extrusion plastometer. A heated barrel holds the thermoplastic, a piston transfers a specified load to the melt, and a die defines the outlet geometry. The mass method collects and weighs the extrudate from a timed interval. The displacement method observes the movement of the piston during a valid part of its travel to calculate the volume displaced. Test conditions define the measured response in both cases.

Conceptual vertical plastometer cross-section shows the applied mass on a piston, a heated barrel holding molten polymer, a specified die and strands of extrudate emerging below.
Generic extrusion plastometer diagram. The applied test load drives the piston downward and pushes molten polymer through the specified die. It does not identify a particular Pacorr model or define a test configuration.

Suitable specimen forms can include pellets, granules or prepared pieces, depending on the material and method. A representative specimen matters because a small charge from a poorly mixed lot may omit unusual pellets, filler variation or another portion of the feedstock. This article focuses on calculating, checking and documenting the result. For an apparatus overview and applications, see Pacorr's Melt Flow Index Tester working-principle guide.

What is the difference between MFI, MFR and MVR?

MFI, short for melt flow index, is the everyday term used in purchase specifications and production discussions. It often refers to the mass-based result. MFR means melt mass-flow rate and is reported in grams per ten minutes. MVR means melt volume-flow rate and is reported in cubic centimetres per ten minutes. Read the unit and procedure from the original record when a display, data sheet or certificate says only flow or index.

Term What to record Common unit
MFI Shop-floor or supplier shorthand. Identify the actual property on the test report. Often g/10 min when it means MFR.
MFR Mass flow, measured directly from timed cuts or calculated from MVR with valid melt density. g/10 min
MVR Volume flow, determined from the permitted displacement process or calculated from MFR with valid melt density. cm³/10 min

Avoid dropping the units when copying a result. The same numerical value of 7.30 can represent mass flow or volume flow, depending on the unit. Use the measurement record and identify how the result was obtained.

How is the MFI formula derived?

In a direct mass determination, divide the measured extrudate mass by the collection time. With grams and seconds, m ÷ t is the flow rate in g/s. Ten minutes contains 600 seconds, so multiply by 600. For several accepted cut-offs of equal duration, use the mean accepted mass with the duration of one cut according to the method. The symbols in this article describe the arithmetic rather than a model-specific display.

Direct mass measurement gives MFR equals 600 times the measured extrudate mass in grams divided by the matched interval in seconds, yielding grams per 10 minutes.
For one valid timed cut, use its mass in grams and its interval in seconds. The factor 600 converts a per-second rate to a per-ten-minute reporting basis.

The time input must represent the mass input. A spreadsheet that uses total mass from several cuts alongside the duration of one cut exaggerates the result. Conversely, mean mass divided by total time understates it. For equal intervals and valid segments, total mass divided by the sum of their durations is arithmetically equivalent to mean mass divided by one duration. The method determines which cuts qualify and which aggregation to report. Do not mix conditions or introduce a new averaging rule when reconciling a lab result. ISO 1133-1:2022.

How do you calculate an MFI of 7.30 g/10 min?

Use a fictional valid extrudate cut with a mass of 0.365 g collected over 30.0 s. Divide 0.365 by 30.0 to find about 0.01217 g/s. Multiply the unrounded rate by 600 to obtain 7.30 g/10 min. These are example inputs, with no prescribed cut interval, target grade or report precision implied.

Illustrative timed cut of 0.365 g over 30.0 s gives 600 times 0.365 divided by 30.0, or 7.30 grams per 10 minutes.
This arithmetic example uses 0.365 g over 30.0 seconds to obtain 7.30 g/10 min. The values are illustrative. Use the timing, cut acceptance and rounding rules of the selected method.

If the balance exports 365 mg, convert to 0.365 g before using the formula. A time export in minutes requires equal care. For this example, 30.0 s equals 0.500 min, which gives 0.730 g/min. Multiply that per-minute rate by 10 to obtain the same 7.30 g/10 min. Use one unit system in a worksheet and label every input field with its unit. Keep unrounded values until the method asks for rounding.

How do you check a worksheet with several cuts?

Suppose a method permits three accepted cuts of equal duration and their masses are 0.361 g, 0.365 g and 0.369 g. The mean is 0.365 g. With 30.0 s per cut, the individual results are 7.22, 7.30 and 7.38 g/10 min, and the mean is 7.30 g/10 min. The total mass is 1.095 g over 90.0 s, yielding the same arithmetic result. The cut count and timing are illustrative. This check only works as shown for cuts taken and combined in accordance with the method.

Review cut order as well as the average. Large cut-to-cut drift can indicate a time-dependent sample, an irregular cut, interruption, or material entering the wrong measurement window. Compare the underlying records and follow the method's criteria for validity and repeat testing. A smooth set of numbers still needs valid preparation and apparatus checks. Keep the observed masses and durations so the review can be repeated independently.

How do you calculate MVR from piston travel?

In a permitted displacement determination, the volume displaced in the measurement interval is the effective cross-sectional area defined for the cylinder and piston geometry multiplied by the piston travel. With area A in cm², travel L in cm and elapsed time t in seconds, MVR = 600 × A × L ÷ t in cm³/10 min. Use the geometry factor required for the verified setup. The small die opening controls the flow, while the area used in this calculation belongs to the cylinder and piston geometry.

Piston travel L and the effective barrel bore area A give displaced volume A times L. With illustrative inputs 0.720 square centimetres, 1.25 centimetres and 45.0 seconds, MVR is 12.0 cubic centimetres per 10 minutes.
For A in cm², L in cm and t in seconds, MVR = 600 × A × L ÷ t. A represents the effective barrel bore area rather than the die opening. The numerical inputs illustrate the arithmetic only.

For an independent illustrative example, let A = 0.720 cm², L = 1.25 cm and t = 45.0 s. The displacement volume is 0.900 cm³ and the rate is 12.0 cm³/10 min. A raw travel reading of 12.5 mm must become 1.25 cm with this formula. These dimensions and times illustrate arithmetic only. A real test uses the prescribed detection zone, geometry, sampling rule and permitted timing or travel settings.

When can you convert MVR to MFR using melt density?

Use MFR = MVR × ρ when the method permits conversion and ρ is the valid melt density in g/cm³ for the material at the applicable test conditions. The reverse calculation is MVR = MFR ÷ ρ. The units cancel as expected because each cubic centimetre of melt corresponds to ρ grams. Record the origin of the density and the conversion method. A room-temperature value for solid granules, a bulk density of pellets or a data-sheet value at an unrelated condition cannot substitute for the required melt density.

In a separate calculation, MVR of 10.0 cubic centimetres per 10 minutes multiplied by melt density of 0.760 grams per cubic centimetre gives MFR of 7.60 grams per 10 minutes.
A separate density-conversion example. Use the melt density for the material at the test conditions. A room-temperature solid density does not supply this conversion factor.

In a separate example, 10.0 cm³/10 min × 0.760 g/cm³ = 7.60 g/10 min. Treat 0.760 g/cm³ as an illustrative input valid for that hypothetical measurement, with no recommended value for a named resin. Ask how to establish melt density at the test temperature and pressure under the governing method. This matters especially when changing formulation, filler level or test conditions. Label converted results so reviewers can tell whether mass was weighed or calculated. ISO 1133-1 and ISO 1133-2.

How do Procedure A and Procedure B differ?

Procedure A uses timed extrudate cuts and mass measurement. The work can be manual or suitably automated. Procedure B follows piston displacement, using elapsed time for a measured travel or travel over an elapsed interval as the applicable method specifies. The displacement data establish volume flow. Calculate a corresponding MFR only with suitable melt-density information and an allowed calculation route. A timer, automatic cutter and balance can automate the mass method without supplying displacement data.

Decision point Procedure A approach Procedure B approach
Primary observation Mass of accepted timed cuts. Piston travel and elapsed time in the valid measurement window.
Result pathway Direct MFR from mass and time. MVR from geometry and travel, with permitted density-based calculation if needed.
Check before testing Balance, cut interval, sample handling and required piston position. Position measurement, timer, geometry factor, measuring window and density source for any conversion.

Use the prescribed procedure rather than selecting by convenience alone. Low or high flow can require a different permitted setup, cut timing or measurement range. Compare the applicable standard and actual equipment capabilities before committing to either route, especially for mixed materials or a workload spanning widely different flow rates.

Which melt flow standards should you use?

Start with the governing material specification or purchase agreement. The current standards below cover related measurements. Their detailed requirements can differ, including apparatus geometry or allowable measurement arrangements, material handling, timing and reporting. Record the edition and method required by the agreement. Results from separate standards need evidence of comparability for the intended decision.

Standard Scope relevant to selection
ASTM D1238-26 Extrusion-plastometer flow rates for molten thermoplastics. Check the selected procedure, material requirements and test condition.
ISO 1133-1:2022 Standard MFR and MVR methods for thermoplastics. Procedure A measures mass and Procedure B measures displacement.
ISO 1133-2:2011 Melt flow for materials sensitive to time-temperature history or moisture. It tightens thermal timing and temperature controls and addresses moisture-sensitive specimen handling.

An ISO Part 2 requirement can change how a lab manages the period from loading to measurement and protects a dried specimen against moisture uptake. Replicating only the set temperature and load does not demonstrate compliance with either standard. Establish the exact observable result, procedure configuration and verification plan before buying equipment or adopting another laboratory's worksheet. ASTM D1238, ISO 1133-1 and ISO 1133-2.

How do you choose temperature, load and die?

The grade, customer specification and test method determine the test condition. Confirm the actual barrel temperature, the complete applied-load configuration and the required die. A pair of familiar numbers from another grade is not a recipe for your sample. A machine's listed weights and temperature limit describe its offered configuration. They do not assign test conditions to a material.

Record die identity, including any authorised alternate geometry. Two dies can give different results with the same resin and load. Use the die specified for the chosen method and the appropriate material, confirm its dimensions and condition, and identify its use in the report. A temperature change can alter viscosity and the time the material spends melting. A different load changes the driving force. If a supplier's result was obtained under different conditions, compare original test records before treating a new number as a lot change.

How do you prepare moisture-sensitive or recycled specimens?

Take a sample that represents the lot and record the form, source, additives or blend details when known. Use the specimen preparation required by the material and method. Dust, residue and mixed pellets can change an individual charge. For recycled feedstock, define the sampling plan and keep isolated findings tied to sampled portions. Flow data can reveal variation while other tests establish identity, contaminants or mechanical suitability. Pacorr's guide to testing recycled plastics.

For a moisture-sensitive material, dry or condition it only as specified for that grade and method, then manage storage and transfer to maintain the intended state. Record dryer settings, exposure time, relevant moisture criteria and any transfer delay according to the controlled procedure. For some polymers hydrolysis can reduce chain length and increase flow under matched conditions. Other changes, including reactions that grow chains or form cross-links, can reduce flow. Diagnosing the mechanism calls for evidence beyond one MFR reading. Keep thermal history visible in the lab record.

What is the melt flow index test procedure?

This educational sequence helps a laboratory plan a repeatable test. Use the controlled standard and instrument instructions for exact specimen mass, temperatures, loads, allowable measurement windows, cleaning and safe handling of hot components.

  1. Identify the grade, lot, required standard, procedure, outcome and test condition. Confirm whether to report MFR, MVR or both.
  2. Prepare and protect a representative specimen. Verify any drying, conditioning, sampling and specimen-form requirements.
  3. Check the die, barrel, piston, temperature measurement, timer and whichever balance or displacement system the test uses. Verify the required load and the agreed instrument checks.
  4. Stabilise the apparatus. Charge the barrel, control the preheat and loading events, and apply the specified load as directed. Observe the method's start-of-measurement rule.
  5. Measure in the valid piston window. Collect and weigh timed cuts for the mass route or acquire the prescribed travel and time data for the volume route. Document unusual flow or instrument events.
  6. Review valid readings and calculate or verify the result. Apply the named rounding, replicate and reporting rules, including the source of melt density for any conversion.
  7. Record results with the test condition and clean the relevant components according to the instrument procedure before running the next material.
Six-step educational sequence: choose the governing method, check the equipment, prepare the sample, charge and preheat by the method, capture mass or displacement, and record the results and conditions.
Educational workflow for a melt flow test. Follow the controlled method and the instrument instructions for hot surfaces, test settings, handling and cleaning.

Use a planned handoff between the operator and instrument software. A computer may store the time window, aggregate cuts or calculate MFR automatically. Check the run configuration and verify an export against the underlying observations before relying on the summary. The charge, preheat and measurement clocks can refer to different events. Align the worksheet timestamps with the test method's definitions.

Why do two labs get different MFI results?

Compare underlying records before adjusting equipment or disputing a lot. Confirm grade and sample identity, the procedure and edition, temperature, applied load, die and melt density if relevant. Then compare the timing from charge to start, conditioning history and the observation or cut window. Duplicate runs from a controlled sample can help distinguish an instrument difference from sample variation. Share raw mass, displacement and time data where possible. Pacorr's guide to different melt flow results between laboratories.

Observation First checks
All results shift after changing software. Check output units, time units, stored cut rule and selected procedure.
Successive cuts trend up or down. Review preheat, timing, moisture, thermal stability and the valid piston zone.
Flow drops after testing a different grade. Inspect the die for residue or partial obstruction and review the approved cleaning record.
MFR and MVR estimates disagree. Check matching specimen conditions, melt-density source and the two raw channels.

Distinguish repeated cuts from repeated charges. Cuts in one run may share a specimen or timing problem. Independently prepared charges can help reveal sampling or preparation variation. Follow the quality plan when invalid runs, deviations or a failed instrument check affect earlier data.

What does a high or low MFR mean in production?

A higher MFR indicates greater mass flow under the same valid conditions. Use it to investigate whether lots of the same grade respond differently or whether a process change coincides with a measured shift. The useful range depends on the product specification and application. A higher number is not a general guarantee of strength, toughness, mouldability or quality. Materials with similar MFR can differ in density, filler, moisture response and performance after forming.

MFR is a condition-dependent test result. It does not directly measure viscosity in Pa·s, identify molecular weight distribution or predict the behaviour of every production line. When a project needs a viscosity curve at processing-relevant shear rates, use an appropriate rheological method. When the concern is mechanical strength, choose a test for that property, such as Pacorr's tensile testing machine range. ASTM D1238 and ISO 11443:2021.

What is the flow rate ratio?

Flow rate ratio, often abbreviated FRR, divides results from specified conditions in a stated order. For example, a defined numerator of 21.0 g/10 min divided by a denominator of 7.00 g/10 min gives 3.00. The illustrative rates use the same result type and units. State the two conditions, their order, the die and the procedure so a reader can reproduce the ratio. ASTM D1238 describes Procedure D as one route for measurements at more than one condition in a single charge.

A specified ratio can aid grade comparisons, but the value depends on the selected conditions and does not by itself identify the material's molecular weight distribution. Use the ratio only for the intended material specification or comparative study. Keep the individual rates beside it to avoid hiding a shift in both measurements.

What goes on an MFR or MVR report?

Every result needs a quantity, unit and conditions. A line can show MFR = 7.30 g/10 min, followed by the actual prescribed temperature and load, the standard and method details, specimen identification and result source. Keep raw measurements and their timestamps. Record exceptions, repeats and what happened to readings outside the valid test window. Include the density source for a calculated conversion and the rule used to reproduce the summary.

Melt-flow report checklist keeps material identity and preparation, method and procedure, temperature and applied load, and the result quantity, unit, observations and deviations together.
Record the test conditions with each result. Compare values only after checking that the method, temperature, load, die and sample preparation are compatible.
Report field What to preserve
Method and settings Standard edition, procedure, temperature, applied load and die.
Sample and preparation Grade, lot, specimen form, sampling, drying or conditioning, and relevant timing.
Raw data Mass and time or travel and time, valid intervals, and instrument identifiers.
Result MFR or MVR with units, agreed precision, summary and density source where applicable.
Disposition Approved limits, decision rule, deviations and any authorised repeat.

A spreadsheet can check its expected units and reconcile sample IDs before reporting. If a result lies near a product limit, follow the lab's agreed decision rule and the material specification rather than rounding early or silently replacing the original result. Record enough detail to trend data through changes in grade, procedure or instrument configuration.

How do you verify and maintain a Melt Flow Index Tester?

Inspect the barrel, die and piston for residue, wear or damage. Use the machine's approved tools and cleaning procedure to avoid changing the die geometry. Verify the dimensions, condition and identification of the die and piston on the required schedule. Check the temperature system using an appropriate verified thermometer or probe and the positions, time and range prescribed by the method. A controller's display step is a resolution figure, while verification addresses performance in the configured barrel.

Check the specified load configuration, timer and balance with the appropriate traceable references. For displacement-based methods, verify the position measurement and travel/time link and confirm the software uses the appropriate area factor. Keep records of calibration, maintenance and pre-run checks. A suitable check material can monitor consistency between calibrations. Investigate changes using the agreed limits and review potentially affected results if verification fails.

What does Pacorr publish for the PCAMFIC-1?

The supplied Pacorr Melt Flow Index Tester listing and its linked brochure identify the PCAMFIC-1 and describe automatic cutting, weighing and MFR calculation. Published specifications include a temperature range up to 400°C, 0.1°C display resolution, timer range up to 999.9 seconds, and listed 2.16 kg and 5.0 kg weights. Pacorr also lists an RS-232 connection and Windows-based software. Ask for the exact supplied load configuration, die, working flow range, programmable intervals and the evidence supporting temperature performance for your method.

If your specification calls for direct MVR, verify a suitable displacement system, verified measurement window and relevant output in the chosen equipment. The published automatic cutting and weighing features describe the mass workflow. Ask to see a representative run and its export when the test plan relies on direct displacement, density conversion or traceable raw data.

What affects the price of a Melt Flow Index Tester?

The quotation depends on the required temperature and flow range, included dies and weights, balance or displacement capability, automation, data export and verification documentation. Installation, operator training and service coverage can also affect the delivered cost. Pacorr's product page gives general market guidance. Request the price of the exact configuration in a written quotation.

Compare offers against the same workload. State your grades, controlling methods, temperatures, loads, expected results and daily volume. Ask which standard accessories, spare dies, sample-handling tools and calibration or verification records are included. If accepting a machine based on a trial, agree which specimen and test output will establish that the configuration meets your requirement.

Frequently asked questions

What is the MFI formula in g/10 min?

Multiply the valid extrudate mass in grams by 600 and divide by the collection time in seconds. For equal-duration accepted cuts, use the mean mass with one cut interval according to the method.

Does a melt flow test always need a ten-minute collection?

The result is expressed on a ten-minute basis. Collect or measure over the valid interval in the selected method and scale the observed rate to 600 seconds.

What is the MFI of a 0.365 g cut collected in 30.0 seconds?

The calculation is 600 × 0.365 ÷ 30.0 = 7.30 g/10 min for that illustrative valid cut.

What is the difference between MFR and MVR?

MFR describes melt flow by mass in g/10 min. MVR describes melt flow by volume in cm³/10 min. Record the procedure and the source of any density used to convert the quantities.

Can I convert MVR to MFR using solid polymer density?

Use melt density for the material at the applicable test conditions and only where the method permits conversion. Normal solid density or bulk pellet density cannot substitute for that value.

Does higher MFI mean a better polymer?

A higher MFI means higher mass flow under the stated conditions when MFI denotes MFR. Suitability depends on the grade specification, processing needs and required finished-product properties.

Can I compare results from different temperatures or loads?

Treat them as distinct test conditions. Compare against the same specified condition for acceptance or use a defined ratio or comparative study when the governing method requires one.

How do I choose between ISO 1133-1 and ISO 1133-2?

Use the method applicable to the material and agreed specification. ISO 1133-2 addresses thermoplastics with significant sensitivity to time-temperature history or moisture during the test.

Does automatic cutting and weighing provide direct MVR?

Direct MVR requires suitable piston-displacement measurement and associated verification. Automatic cutting and weighing support the mass method.

How can Pacorr help configure your melt flow test?

Share your grades, standard and edition, selected procedures, temperature and load combinations, expected results, sample conditioning and required output. Include throughput and data-export needs, and say whether you require direct displacement or a density-based conversion. Contact Pacorr to discuss the equipment configuration and request a quotation.

Our USP's

01

Demonstration of Machine

  • Try Before You Decide
  • Experience Before You Invest
  • See It in Action – Free Demo
02

Global Reach

  • Delivering Excellence Worldwide
  • Serving Across Borders
  • Your Global Supply Partner
03

On-Site Support

  • Support That Comes to You
  • On-Ground Assistance, Anywhere
  • We Reach Where You Are
04

Team Strength

  • Powered by Expert Minds
  • Driven by Skilled Professionals
  • Engineering Excellence Team
05

Manufacturing

  • Precision-Driven Manufacturing
  • Smart & Automated Production
  • Engineered In-House Excellence
Testimonials

What Our Customers Say

Thanks to Pacorr Testing instruments, we have all the required quality testing instruments that have helped us to ensure the best quality delivered to our clients.

D

Danish

Fair Exports Pvt. Ltd.

The extensive range of testing instruments they have to offer, satisfied all my quality testing needs. The most impressive things, I feel, is the comprehensive after-sales support that I received from them.

NJ

Nipun Jain

Fast Pack India

I recently ordered some plastic molding testing instruments from Pacorr and I must say I am amazed by the quality of their devices and the support we got from their team was remarkable. I totally recommend Pacorr instruments for unmatched quality assurance.

BS

Bhaskar Srivastava

Dipty Lal Judge Mal Pvt Ltd

Pacorr has offered us the best range of chemicals and coating testing instruments that have made it extremely easy to take care of our product quality. Thanks Pacorr Testing Instruments.

RD

Rajpal Dhiman

Tasmay Unicoats Private Limited

I purchased a torque tester for my company which is working just perfectly fine since last 6 months. Also, their comprehensive after sales service makes them stand out from others. I will recommend Pacorr for best quality products and excellent customer service.

AJ

Mr. Anuj Jakhodia (Director)

Safe Closures LLP

WhatsApp