Portable Spectrophotometer: Delta E Formula for Colour Matching

A Portable Spectrophotometer measures how a specimen reflects light across wavelengths. In colour matching, the instrument or its software uses those measurements to calculate colour coordinates under defined conditions. Delta E expresses the size of the difference between a sample and a reference. For the CIE 1976 formula, commonly called CIE76, **ΔE*ab = √[(ΔL*)² + (Δa*)² + (Δb*)²]. With a reference of (50, 10, 20) and a sample of (53, 14, 20), the directional differences are +3, +4 and 0, and ΔE*ab = 5.00**.

The result depends on the calculation and the measurements that feed it. Use the same illuminant, observer, optical geometry and specimen presentation for both readings. Name the formula with the result because CIE76, CIE94 and CIEDE2000 can return different values for the same coordinate pair. The worked examples in this guide explain calculation and interpretation. Use the agreed product method, instrument instructions and acceptance rules for actual decisions.

What does a portable spectrophotometer measure?

A reflective colour instrument illuminates a defined spot and records a spectral reflectance factor relative to its reference under the specified geometry. The resulting spectrum retains information about how the sample responds at different wavelengths. To calculate object colour, software combines spectral data with the chosen illuminant and observer functions to obtain tristimulus values X, Y and Z. It transforms those values relative to the corresponding reference white to calculate CIELAB coordinates.

The route from spectrum to coordinates matters when two instruments produce different results. The recorded spectrum depends on measurement conditions and sample presentation. The calculated coordinates also depend on the illuminant, observer and reference white. Retain spectral data where practical so a lab can audit the calculation or evaluate an additional permitted illuminant. For a wider explanation of optical instrument types and applications, see Pacorr's spectrophotometer working-principle guide. CIE 015:2018, Colorimetry.

What do L*, a* and b* mean?

CIELAB places a colour stimulus at a point described by three coordinates. L* represents lightness. The a* and b* coordinates describe opponent colour directions. When the article subtracts reference from sample, the sign tells you which way the sample moved along that axis.

Three directional scales show L* from darker to lighter, a* from negative greenward to positive redward, and b* from negative blueward to positive yellowward.
CIELAB coordinates describe lightness and positions on two opponent-colour axes. The graphic is a directional guide with schematic colours.
Coordinate Meaning Positive sample minus reference
L* Lightness. Sample is lighter on the L* scale.
a* Green to red direction. Shift towards red, or away from green.
b* Blue to yellow direction. Shift towards yellow, or away from blue.

An increase in a* can move a negative reading closer to zero while it remains on the green side of the axis. Similarly, an increase in b* can make a negative b* less blue without making it positive. Keep the signs and actual coordinates, especially when the lab uses them to investigate shade drift. ISO/CIE 11664-4:2019.

How do you calculate the CIE76 Delta E formula?

Calculate **ΔL* = L*sample − L*reference**, and apply the same subtraction order to a* and b*. Square the three differences, add the squares and take the square root. The formula gives the straight-line distance between the two points in CIELAB. It produces a nonnegative number and has no percent or physical unit. Keep enough coordinate precision to perform the calculation before rounding the result for a report.

CIE76 Delta E is the square root of the sum of squared Delta L*, Delta a* and Delta b*. Each signed component subtracts the reference coordinate from the sample coordinate.
CIE76 calculates the Euclidean distance between two CIELAB measurements. Retain the signed differences for direction and identify the formula used when reporting Delta E.

Reversing the two readings reverses the signs of the three directional differences while leaving the CIE76 distance the same. In practice, reserve one order for all spreadsheet and report fields. Label which reading was the reference so a second person can interpret the differences correctly. Separate colour-difference formulas can use a reference-dependent weighting, so maintaining the correct role also matters when the selected method uses those formulas.

How does the 5.00 Delta E calculation work?

Suppose the reference has L* = 50, a* = 10 and b* = 20. The sample has L* = 53, a* = 14 and b* = 20 under the same measurement conditions. Then ΔL* = +3, Δa* = +4 and Δb* = 0. Calculate √(9 + 16 + 0) = √25 = 5.00 using CIE76. In this example the sample is lighter and shifted towards red. The b* coordinate did not change.

CIE76 example: Reference L*, a*, b* are 50, 10, 20 and the sample values are 53, 14, 20. Sample-minus-reference differences are +3, +4 and 0, so Delta E is 5.00.
Illustrative measurements give ΔL* = +3, Δa* = +4 and Δb* = 0. The CIE76 calculation gives ΔE*ab = 5.00 under the assumed matching measurement conditions.

One overall value can describe many different coordinate shifts. Another hypothetical sample could have ΔL* = −5, Δa* = 0 and Δb* = 0 and still give 5.00 using CIE76. Its change is entirely towards lower lightness. Record the directional differences alongside the overall value to investigate whether a change is dominated by lightness or one of the opponent axes. The numbers are arithmetic examples, independent of any tolerance or product grade.

How do you distinguish the axis shifts from chroma and hue?

The a* and b* changes describe movement along axes. Chroma describes the radial distance from the neutral point in that plane. In CIELAB, **C*ab = √[(a*)² + (b*)²]**. In the main example, the reference chroma is about 22.36 and the sample chroma is about 24.41. Chroma increases by about 2.05 with Δa* = +4 and Δb* = 0. An unchanged b* coordinate does not hold hue constant because hue depends on the direction of the point from the origin in the a*, b* plane.

When a customer requests a lightness, chroma and hue breakdown, follow its method and use the corresponding definitions. A hue angle difference in degrees and the hue term used by a colour-difference formula describe distinct quantities. Treat hue near neutral with care and use the prescribed calculation for angular wraparound. The plain CIELAB chroma example above uses no CIEDE2000 corrections.

Why do CIE76, CIE94 and CIEDE2000 differ?

The three calculations use CIELAB information differently. CIELAB is approximately uniform and equal straight-line distances do not always correspond to equal perceived differences. Weighted formulas adjust the contributions of lightness, chroma and hue to improve agreement with relevant visual data. Adopt the named method for the product and keep it the same when comparing results or applying historical limits.

Formula What the calculation does
CIE76, ΔE*ab Euclidean distance in the L*, a*, b* coordinates.
CIE94, ΔE94 Weights lightness, chroma and hue differences using reference and application parameters.
CIEDE2000, ΔE00 Uses adjusted lightness, chroma and hue terms with weighting and a chroma–hue interaction term.
CMC, ΔEcmc Weights components relative to the reference colour. Record the specified lightness-to-chroma ratio.

CIE94 application parameters used for graphics and textiles can differ. A report stating only ΔE94 leaves out information when the setting affects the calculation. CMC uses a ratio such as 2:1 or 1:1 when specified. Where parameters can be changed, record what the software actually used. A result obtained with one formula cannot be converted to another from the scalar number alone. CIE 116-1995 and ASTM D2244-25.

How does CIEDE2000 calculate colour difference?

ISO/CIE 11664-6:2022 defines CIEDE2000 for CIELAB input values calculated under the applicable conditions. The calculation adjusts the a* coordinate for chroma behaviour and derives differences in lightness, adjusted chroma and adjusted hue. It scales those components using their weighting functions and the selected parametric factors. An interaction term combines the scaled chroma and hue terms and matters particularly in the blue region. These operations use both colour points. The three signed axis differences alone do not contain all the starting information for ΔE00.

For a separate software check, the reference test pair (50.0000, 2.6772, −79.7751) and (50.0000, 0.0000, −82.7485) in Sharma, Wu and Dalal's supplementary dataset gives ΔE00 = 2.0425 with unit parametric factors. This pair is unrelated to the CIE76 example above. Use several published test pairs when verifying a spreadsheet or application because a single pair may miss incorrect hue wrapping, angle units or handling of near-neutral colours. ISO/CIE 11664-6:2022 and Sharma, Wu and Dalal's implementation test data.

Which measurement conditions must match?

Copy the complete measurement configuration from the approved method before reading the reference and sample. Match the settings and presentation, then retain them in the record. A stored target from another laboratory is usable only if its conditions and equivalence are established for the comparison.

A generic 45-degree/0-degree cross-section shows illumination at 45 degrees from the surface normal and observation along the normal. A checklist covers aperture, backing, orientation, illuminant and observer.
Illustrative 45°/0° side view. The angles describe illumination and observation relative to the surface normal. Match the required geometry and other measurement conditions for both readings.
Condition What to match or record
Colour calculation Colour space, named Delta E formula and applicable parameter settings.
Illuminant and observer Same illuminant and same standard observer for both coordinate sets.
Optical arrangement Measurement geometry, aperture, mode, specular treatment where applicable and UV setting where specified.
Specimen presentation Backing, layer count, surface, orientation, measurement locations and relevant conditioning.

A larger or smaller aperture can include a different portion of a patterned sample. A different observer changes the colour-matching functions used in the calculation. A different mode can change the recorded spectrum. Reassess the raw measurement and settings when a comparison crosses one of these boundaries.

What are illuminants and standard observers?

An illuminant specifies how optical power is distributed by wavelength for the colour calculation. For example, CIE D65 represents a daylight spectral distribution and CIE A represents typical incandescent illumination. An instrument may use its own physical light source while calculating object colour for a specified illuminant with the measured spectrum. Do not assume that selecting D65 on a screen describes the physical illumination used on a visual inspection bench.

The CIE 1931 2° and CIE 1964 10° standard observers provide different sets of colour-matching functions. These angles relate to the visual field represented by the functions. They do not describe the instrument's illumination or detector geometry. A label such as D65/10° therefore specifies two calculation choices, with the physical measurement geometry recorded separately. CIE standard illuminants and CIE standard colorimetric observers.

How do measurement geometry and gloss affect colour readings?

Geometry defines the angles or distribution of illumination and collection relative to the sample. A conventional 45°/0° arrangement illuminates at approximately 45° to the normal and views along the normal. In instruments offering integrating-sphere modes, specular-included and specular-excluded settings have distinct collection conditions. Confirm the precise geometry, directionality and available modes from the instrument documentation. Match both reference and sample when comparing colour.

Surface gloss, texture and direction can influence the relationship between instrument values and visual appearance. If a product also specifies gloss in gloss units, use the prescribed gloss method and geometry. A colour-difference value alone cannot establish gloss conformity. For a separate explanation of standard gloss angles, see Pacorr's gloss meter measurement guide.

Why do aperture, backing and orientation matter?

Use an aperture and specimen that allow the prescribed region to be measured without including an edge, hole or different colour zone unless the protocol requires it. Locate and mark regions where practical. Check the head sits flush, especially on curved mouldings or small parts. A gap can let in ambient light. A curved or metallic-effect surface may require a specialised fixture or angular measurement depending on the specification.

Thin, translucent or partially opaque specimens can pick up colour from whatever lies behind them. Use the specified backing, thickness or layer arrangement. A textile with pile or a directional finish can change with rotation. Define the orientation or approved rotations before the run. Keep printed ink, gloss level, coating thickness, cure, substrate and weave traceable when any could influence the result.

How do you measure a sample against a reference?

Use this educational workflow to plan comparable measurements. Obtain the exact preparation, instrument checks, settings and replicate rules from the agreed method and model instructions.

  1. Identify the physical or verified stored reference, the sample lot and the surfaces to compare. Record the formula, illuminant, observer, geometry, aperture and applicable modes.
  2. Prepare representative areas. Apply the specified backing, condition and orientation. Remove unintended contamination using the documented procedure.
  3. Standardise and verify the instrument with the required references under the manufacturer's instructions and the laboratory procedure.
  4. Measure the reference and sample at the defined locations and orientations. Reposition the instrument for replicate readings as required, and retain individual data.
  5. Calculate the signed component differences and selected Delta E. Check units, the reference identity, calculation method and any application parameters before exporting.
  6. Review variation, repeat checks and visual findings as applicable. Apply the agreed acceptance rule and preserve the results and deviations.
Six-step measurement sequence identifies the method, prepares and verifies the instrument, prepares the surfaces, measures the reference, measures the sample in matching conditions, and records Delta E with its formula and settings.
Educational measurement sequence. Follow the instrument's instructions and your procedure for calibration, reference checks, repeat measurements and result reporting.

If software offers to use a stored reference automatically, check its revision and origin before running a new batch. A replacement master with a similar appearance can have different coordinates. A reference swap must remain traceable for production trend data to be meaningful.

What is the difference between standardisation and verification?

Follow the instrument procedure for white-reference and black-reference or zero steps where applicable. These steps establish the instrument's working response for its measurement configuration. Protect measurement references from damage, contamination and unintended light exposure. A reference used to adjust the instrument serves a different role from a suitable independent check material used to confirm performance.

Track the reference identifiers, condition, check results and service history. Follow the lab plan for routine checks and after changes that can affect results. A small spread from repeated white-tile readings can confirm one part of instrument performance. It does not represent lot variation, placement effects or the overall uncertainty of a measurement on a textured part. Investigate a failed check before using a new result for product acceptance.

How do you average replicate colour measurements?

Use the summary required by the method. Averaging the sample's L*, a* and b* values first and then calculating one Delta E defines a different summary from calculating each Delta E first and averaging those values. For example, two sample points with only ΔL* = +2 and −2 each give CIE76 2.00. Their mean coordinates give a CIE76 distance of zero from the reference. The averages answer different questions about the surface.

Repeated readings without moving the instrument primarily assess short-term indication repeatability at one position. Readings after repositioning include placement effects and can include specimen variation. Keep individual positions and directions, especially if a maximum local difference or several named zones determine acceptance. Repeated measurement of one position cannot establish uniformity across a batch.

What is metamerism and how can you check it?

Two samples may match closely under one illuminant and differ under another if their reflectance spectra differ. This is an example of illuminant metamerism. A formula such as ΔE00 evaluates a difference for the input coordinates. Its result under one specified illuminant does not establish agreement under the other illuminants relevant to the product.

Two schematic materials have different spectral reflectance curves. The pair may match under one illuminant and separate under another. The graphic uses no measured spectra or exact viewing colours.
Conceptual metamerism example. An apparent match under one illuminant can change under another. Evaluate the actual specimens under the required illuminants and viewing conditions.

Where the optical method supports it, compare the spectra and calculate colour differences separately for each required illuminant with consistent observer and measurement conditions. For samples with fluorescence, confirm the required UV excitation and measurement method rather than assuming the same spectral data can always be reweighted. A controlled visual comparison can complement the instrumental result. Use a suitable light source or Pacorr Colour Matching Cabinet under the method's viewing conditions. Pacorr's visual cabinet workflow guide and CIE metamerism terminology.

What Delta E is acceptable for colour matching?

There is no universal limit. The product requirement, visually accepted samples, method, material and customer determine an appropriate tolerance. A limit stated for ΔE00 cannot simply be used for ΔE*ab. Check whether separate lightness or colour-direction limits apply, and whether the requirement concerns each reading, a summary or an agreed set of zones.

Suppose a fictional control plan specifies a maximum ΔE00 of 1.00 and a conservative rule that accepts a result only when the measured value plus the applicable expanded uncertainty is no greater than 1.00. A value of 0.95 with expanded uncertainty 0.10 gives 1.05 under that rule, so it does not meet this example's acceptance criterion. The rule is illustrative. Use the agreed disposition for a result close to a real limit, and keep displayed rounding separate from the data used for the decision. ILAC guidance on decision rules.

Why can two instruments give different Delta E readings?

First compare the measurement configurations and the underlying coordinate data. If both instruments agree on the coordinates but disagree on Delta E, investigate the software formula, parameters and order of inputs. If the coordinates differ, inspect the measured spot, optical settings, reference checks and sample handling before comparing specifications or recalculating.

Observation First check
Same coordinates, different total difference. Check formula, parameter choices, input order and rounding.
Readings shift after turning the sample. Check directional finish, placement, aperture and approved orientations.
Reference agrees while production samples differ. Use comparable regions and check geometry, aperture, backing and surface gloss.
One site's results shift across several colours. Check the instrument's reference history, software configuration and specimen preparation.

Use representative shared standards or parts when establishing whether instruments agree for an application. A catalogue agreement figure obtained on a set of tiles does not tell you their agreement on every coloured textile, film or coating. Specify the comparison conditions and the allowable decision differences before exchanging data between locations.

How should suppliers and buyers share a digital colour target?

Send the method and target together. Identify the master, its revision, the location or area used and the exact configuration that produced its data. Include full-precision coordinates and preserve the measured spectrum when available. Check how software transfers the illuminant, observer, geometry, formula and tolerances. A data file can retain a coordinate value while losing a setting that is needed to interpret it.

During method transfer, compare both systems on shared representative parts or standards. Differences between the instruments on a white tile and differences on the actual construction can have different causes. Agree on specimen placement and orientation, then compare the individual data and the decisions they would produce. Where physical masters are in use, store and handle them under an agreed plan so that ageing, abrasion or replacement does not silently redefine the target.

What belongs in a colour matching report?

Retain enough information for another lab to reproduce the comparison. In the first example, report CIE76 ΔE*ab = 5.00, the two sets of coordinates and their signed differences. Add the actual instrument conditions and the rule that governs the decision. Keep separate results under different illuminants, formula choices or viewing geometries.

Colour difference report checklist includes the reference and sample IDs and surface side, measurement geometry and aperture, illuminant and observer, individual CIELAB values, signed differences, Delta E formula and agreed tolerance.
Report the two readings and the named colour-difference formula with the measurement conditions. Compare against the tolerance agreed for that formula and application.
Report field What to preserve
Specification and target Controlled method, reference ID and revision, sample and lot, and reporting requirements.
Measurement conditions Instrument ID, geometry, aperture, illuminant, observer, modes and specimen presentation.
Results and checks Raw coordinates and spectra where available, replicate locations, signed differences, named Delta E formula and verification records.
Decision Applicable limit, uncertainty or decision rule where required, deviations and disposition.

For a colour issue in production, save any separate gloss or visual observations with their own method. The report should make it possible to see whether an apparent mismatch follows a process change, an instrument condition or a difference in the locations measured.

What does Pacorr publish for the NS810-PC?

The Pacorr Portable Spectrophotometer listing identifies NS810-PC in the model field and refers to NS810 in the technical overview. It lists a 45°/0° illumination/viewing arrangement, 8 mm measuring aperture and 400–700 nm spectral range at 10 nm intervals. The listed calculation options include 2° and 10° standard observers and differences in CIE76, CIE94, CIEDE2000 and the stated CMC ratios. The stated measurement time is 1.5 s.

Confirm the exact supplied model, physical geometry, available modes and export fields in the quotation and demonstration. Ask for representative data from your actual material if the requirement involves a narrow area, translucent film, fluorescent substrate, highly directional finish or additional appearance measurement. Consider whether verification materials, software and test data formats match the quality plan.

What affects the price of a portable spectrophotometer?

Price depends on the optical arrangement, available apertures, specimen accessories, software, reference materials, verification documentation and service requirements. Pacorr's listing gives a general price indication that varies with configuration. Request a written quote for the instrument and accessories required by your method. Ask whether a trial on real parts, training or ongoing service is included.

Frequently asked questions

What is the Delta E formula in CIELAB?

For CIE76, calculate the square root of the sum of the squared differences in L*, a* and b* between sample and reference measured under matched conditions.

Is Delta E a percentage?

Delta E is a numerical colour-difference magnitude for the named formula. It is reported without a percent sign.

What does a positive Δa* mean?

With sample minus reference subtraction, a positive Δa* is a change towards red or away from green relative to the reference.

Is ΔE00 the same as ΔE*ab?

They use the same CIELAB input space, but ΔE00 uses CIEDE2000 weighting and interaction terms. ΔE*ab uses the CIE76 straight-line distance.

Can I compare samples measured with different illuminants?

For an acceptance comparison, calculate and record both readings under the same specified illuminant. For a metamerism study, make a separate matched comparison under each required illuminant.

Does Delta E of zero guarantee identical appearance?

Zero indicates matching calculated coordinates for the chosen formula and conditions. Spectra, gloss and appearance under a different viewing condition can still differ.

Is Delta E below 1 always acceptable?

Acceptance depends on the named formula, material, customer's limit, measurement conditions and decision rule. Apply the specification agreed for the product.

Can a colour spectrophotometer replace a gloss meter?

Use the method and geometry specified for gloss if gloss units are required. A spectral colour result alone does not establish compliance with a gloss requirement.

Why are repeat readings different after moving the instrument?

Repositioning can sample a different area, orientation or local texture and change how the head sits. Control these factors and retain individual readings according to the sampling plan.

How can Pacorr help configure your colour measurement?

Send the sample material and dimensions, finish, required aperture and geometry, formula, illuminant and observer, expected tolerance and desired output. Include your reference type and whether you also need visual or gloss evaluation. Contact Pacorr to discuss the configuration and request a quotation.

Our USP's

01

Demonstration of Machine

  • Try Before You Decide
  • Experience Before You Invest
  • See It in Action – Free Demo
02

Global Reach

  • Delivering Excellence Worldwide
  • Serving Across Borders
  • Your Global Supply Partner
03

On-Site Support

  • Support That Comes to You
  • On-Ground Assistance, Anywhere
  • We Reach Where You Are
04

Team Strength

  • Powered by Expert Minds
  • Driven by Skilled Professionals
  • Engineering Excellence Team
05

Manufacturing

  • Precision-Driven Manufacturing
  • Smart & Automated Production
  • Engineered In-House Excellence
Testimonials

What Our Customers Say

Thanks to Pacorr Testing instruments, we have all the required quality testing instruments that have helped us to ensure the best quality delivered to our clients.

D

Danish

Fair Exports Pvt. Ltd.

The extensive range of testing instruments they have to offer, satisfied all my quality testing needs. The most impressive things, I feel, is the comprehensive after-sales support that I received from them.

NJ

Nipun Jain

Fast Pack India

I recently ordered some plastic molding testing instruments from Pacorr and I must say I am amazed by the quality of their devices and the support we got from their team was remarkable. I totally recommend Pacorr instruments for unmatched quality assurance.

BS

Bhaskar Srivastava

Dipty Lal Judge Mal Pvt Ltd

Pacorr has offered us the best range of chemicals and coating testing instruments that have made it extremely easy to take care of our product quality. Thanks Pacorr Testing Instruments.

RD

Rajpal Dhiman

Tasmay Unicoats Private Limited

I purchased a torque tester for my company which is working just perfectly fine since last 6 months. Also, their comprehensive after sales service makes them stand out from others. I will recommend Pacorr for best quality products and excellent customer service.

AJ

Mr. Anuj Jakhodia (Director)

Safe Closures LLP

WhatsApp